Not available outside of the UK & Ireland.
General description
Method: photometric 0.5 - 12.0 mg/l O2 Spectroquant®The method is analogous to DIN EN 25813-21.This Spectroquant® Oxygen Cell Test allows the accurate quantification of the oxygen concentration in aqueous samples. Method applied: Oxygen oxidizes manganese(II) to manganese(III). In acidic solution, manganese(III) reacts with Titriplex® II to form a red complex which is determined photometrically (modified Winkler method).The Spectroquant® Cell Tests come with prefilled 16 mm round cells and all the required reagents to perform the analysis according to the instruction leaflet provided. All Spectroquant® Cell and Reagent Test Kits are equipped with the unique Live ID (2D barcode) which allows seamless method recognition and contains essential information such as lot number, expiry date, and automatic calibration updates.
Legal Information
SPECTROQUANT is a registered trademark of Merck KGaA, Darmstadt, Germany
Titriplex is a registered trademark of Merck KGaA, Darmstadt, Germany
This product has met the following criteria: