Increase the accuracy of your readings with the Thermo Scientific™ Orion™ ROSS Ultra™ low maintenance pH/ATC Triode™ combination electrodes. These combination electrodes have a pH sensor that delivers superior performance at varying sample temperatures. They combine the stability, precision and speed of response of the ROSS Ultra line with integrated temperature measurement. Orion™ ROSS™ electrodes contain no silver or silver chloride and incorporate a double-junction design, which protects the sample from contamination and makes the electrodes compatible with TRIS buffers.
Low ionic strength samples, such as clean waters, present a challenge to pH electrode performance. Sluggish response, poor accuracy and drift are common. The Thermo Scientific Orion ROSS Ultra 8107 Triode pH electrode responds within one minute, proving a stable, accurate reading. Complete your pH measurements faster. Get long, maintenance-free electrode lifetime with special long-life gel and innovative capillary glass junction design. Maintain sample integrity, even in TRIS buffers; ROSS double junction design electrodes contain no silver, which can bind with biological samples. With built-in temperature sensors, these electrodes offer a convenient solution for your routine sample measurements. Compatible with Thermo Scientific™ Orion™ Versa Star™, Thermo Scientific™ Orion Star™ A, and Thermo Scientific™ Orion™ Dual Star™ meters, as well as Mettler Toledo™ pH meters.
- Connector types: BNC waterproof, 8-pin MiniDIN
- Junction: glass capillary
- Material: rugged epoxy body, filled with low maintenance gel
- pH range: 0 to 14
- Operating temperature: 0 to 80°C
- Type: pH/ATC combination electrode
- Filling solution type: none required
- Internal reference: ROSS
- Isopotential point: 7
- pH relative accuracy: 0.03 pH for buffers using automatic temperature compensation
- Precision: 0.01
- Sample compatibility: TRIS, sulfides, or proteins
- Dims, length x diam: 120 x 12mm
- Length cable: 1m
This product has met the following criteria: